Model S100
FEATURES
The S100 is a basic wide-range spectrometer which is hard to manage without in any scientific laboratory, for spectroscopy education at colleges and universities, and even in field conditions.
EXTREMELY WIDE RANGE – 190 TO 1100 NM – COVERED WITH THE USE OF ONE DIFFRACTION GRATING
COMBINATION OF COMPACT DESIGN AND HIGH RESOLUTION
USER-FRIENDLY INTERFACE COMPATIBLE WITH WINDOWS 2000/XP/7/8/10
NO EXTERNAL POWER REQUIREMENTS
DESCRIPTION
Use of original concave diffraction grating makes the S100 a unique device which provides high resolution of 1nm within extremely wide spectral range from 190 to 1100 nm.
The S100 spectrometer is convenient and easy to use: it is calibrated by the manufacturer, does not contain any movable parts, is controlled and powered from the computer via Full-Speed USB interface.
The S100 can be triggered from your light source with standard TTL trigger pulses via connector BNC-58. The S100 is also able to produce TTL trigger pulses.
The S100 complete set contains USB control cable, direct and reverse synchronisation cables and UV optical fiber.
The spectrometer is able to operate without the optical fiber with analysed light directly steered to its input slit.
Upon your request the S100 may be supplied with an order sorting filter separating high spectrum orders.
Specific features of the S100 spectrometer depend on the type of the selected detection system.
Toshiba linear image sensors TCD1304 and TCD1205 installed in the S100-3648T and S100-2048T feature high sensitivity; large quantity of relatively narrow pixels ensures high resolution and accurate wavelength determination. However, when recording wide-band spectra with the S100-3648T and S100-2048T, spectrum modulation is observed, thus limiting to some extent their application in spectrometry.
The software allows the user to perform calibration of the S100 spectral sensitivity in a selected spectral range. Factory calibration of the S100 spectral sensitivity is not provided.
The S13496-4096 and S11639 -2048 Hamamatsu image sensors installed in the S100 demonstrate excellent linearity, wide dynamic range and absence of spectrum modulation.
APPLICATIONS
Laser wavelength control
Plasma control
Analysis of any light source in the 190-1100 nm spectral range
SPECIFICATIONS
SPECTROMETER MODEL | S100-3648T | S100-2048T | S100-4096H | S100-2048H |
Grating, lines/mm (average) | 300 | |||
Reciprocal dispersion (average), nm/mm | 33.28 | |||
F/Number | 1 : 6 | |||
Focal Length, mm | 99 | |||
Spectral range, nm | 190 — 1100 | 200 — 1100 | 190 — 1000 | 200 — 1000 |
Spectral Resolution (average), nm | 1.0 | 1.5 | 1.0 | 1.5 |
LINEAR IMAGE SENSOR | CCD TCD1304 Toshiba3648 pixels | CCD TCD1205 Toshiba2048 pixels | CMOS S13496 Hamamatsu4096 pixels | CMOS S11639 Hamamatsu
2048 pixels |
Pixel size, µm | 8 x 200 | 14 x 200 | 7 x 200 | 14 x 200 |
Active area, mm | 29.184 | 28.672 | 28.672 | 28.672 |
Minimal exposure time, ms | 7.4 | 4.2 | 0.018 | 0.018 |
Maximal exposure time, not less, s1) | 3 | 4 | 90 | 90 |
Antiblooming2) | NO | Yes | Yes | Yes |
Photo response non-uniformity,%3) | ± 2 % | ± 2 % | ± 5 % | ± 5 % |
Dynamic Range | 2000 : 1 | 1100 : 1 | 5000:1 | 5000:1 |
Photo sensitivity, V/lx×s | 160 | 80 | 650 | 1300 |
Mean-square reading noise, counts | <8 | <14 | < 4 | < 4 |
Data rate (max), kHz | 500 | 500 | 3000 | 3000 |
ADC Resolution | 14 bit, 16384 counts | |||
Triggering | internal / external | |||
Requirements to external trigger pulse | BNC-58 connector, positive polarity, 3-15 VDC amplitude,5-20 µs pulse duration FWHM | |||
Parameters of the S100 trigger pulses | positive polarity, 4-5 VDC amplitude, 10 µs pulse duration FWHM | |||
Computer interface | Full-Speed USB | |||
Optical Input | — Direct input through the S100’s entrance slit
— Optical Fiber : 0.6 (0.4)mm diameter, 1m length, SMA-905 connector |
|||
Overall Size, mm | 66 × 86 × 146 | |||
Weight, kg | 0.75 |
OPTIONS & ACCESSORIES
Upon Your request, the S100 spectrometer can be equipped with high-order separating compound filter designed for operation with broadband spectra.
The S100 spectrometer can be supplied with VUV or FWF optical fibers with various lengths.
The S100 spectrometer has a wide range of options and accessories that allow efficient collection of light from the sample located at a finite distance and in infinity:
Condenser PS-2 is designed for collecting light from a sample located at 40 mm distance within a wide spectral range and its effective transmission into an optical fiber or onto an entrance slit of a spectral device.
Condenser PS-3 is designed for collecting light from a sample located at 29 mm distance within a wide aperture angle and its further transmission into an optical fiber or onto an entrance slit of a spectral device.
Adapter FA-1 – achromatic adapter for focusing parallel beam to a fiber with 3D-alignment.
Adapter FA-2 – achromatic adapter for focusing parallel beam to a fiber with an axial alignment.
Cassegrain objectives allow collecting light from objects located at a distance of 3.5m to infinity.
Diffuse attenuator FA-3 is designed to attenuate the laser beam while delivering it to a fiber.
The S100 spectrometer can be supplied with a cuvette holder FA-10 and with fiber-optics immersion probes DipProbes with a set of interchangeable tips for liquid handling.
Contact a SOLAR LS specialist to select the complete set of the device that meets the requirement of your instrument complex in the best way.
Mercury spectrum acquired with the S100-3648T (order-separating filter not installed).
Deuterium lamp spectrum acquired with the detector CMOS S13496 Hamamatsu (exposure time 500 ms/ 1000 ms/ 3000 ms).
Deuterium lamp spectrum acquired with the S100-3648T (exposure time 500 ms/ 1000 ms/ 3000 ms).