IRA-3-10 Autocorrelator
Pulse duration range 30 fs – 50 ps
Wavelength range 3-10 um
Scan range 150 ps
No reference beam required
USB connection
Acquisition and analysis PC software
1. General
The IRA-3-10 autocorrelator is an indispensable tool for high resolution pulse duration measurement of femtosecond and picosecond OPAs. The unit consists of an optical head and an electronic control unit with USB interface for easy PC control. The software is included in the package and displays the AC trace and calculates pulse duration with Gaussian and sech^2 fitting options and data storage and export.
The wavelength coverage of the unit is divided into three ranges: 3-5.5 um, 5.5-7 um and 7-10 um, each being covered by a relevant exchangeable optics set.
2. Specifications
3. Applications
Femtosecond and picosecond OPA pulse duration measurement.